Set up GAM lift
Connect GAM to GAM motor controller
Navigate to telescope > GAM control > filter wheels
Tried to move the on-axis filter wheel to pos. 3
Connected 37-AMP connector
Moved South filter wheel to 3, SUCCESS
Moved South filter wheel to 6, SUCCESS
Moved South filter wheel back to 3, SUCCESS, and rotated in opposite direction
Moved on axis filter wheel to 3, began to move but wouldn't stop due to missing belt
Moved to the translating stage tab
Rotating mirror to west. Success.
Rotating the mirror works in all directions. Sometimes it gets stuck, can't consistently replicate it, but it corrects itself on the second try. Rotating the mirror by hand shows that it experiences more resistance at some parts of the rotation.
Tried testing the translating stage, moved between stow and on-axis pos. repeatedly, FAILED
The south filter wheel, on-axis filter wheel, translating stage, mirror, all seem to work. The translating stage motor rotates slightly when clicking “Update GAM status”, along with all the rest of the motors. This can be replicated by pressing the reset switch on the side of the GAM motor controller and then clicking the update gam status button on DFMTCS again.